SG-IP Strain Gauge Amplifiers
Single Channel Strain Gauge Amplifier
About the SG-IP Strain Gauge Amplifiers
The single-channel strain gauge measuring amplifier SG-IP allows supply and signal evaluation of one transducer with a strain gauge full bridge greater than 300 ohms. Input and output are galvanically isolated. The connection is done in 4-wire technology. For further evaluation standard analogue outputs are available. The measuring amplifier is built-in into a robust EMC die-cast aluminium enclosure (IP65) which is suited for applications in rough and industrial environments.
The amplification of the amplifier can be adapted by an internal precision resistance.
The potentiometers Z (Zero) and G (Gain) which are accessible after removing the enclosure cover allow a calibration correction. By means of an internal dip switch a change of the range of the potentiometer Z (Zero) can be reached. To allow a zero range shift a basic load / tare can be suppressed by a resistance electrically.
Key features SG-IP
- Supply voltage 10-18 VDC / 18-30 VDC
- Analogue output 0-10 V / ±10 V / 4-20 mA
- EMC Aluminium die-cast enclosure (IP65)
- Dimensions (W x H x D) 175 x 80 x 57 mm
Avez-vous besoin de ce capteur pour votre prochain projet?
Demandez un devis sans engagement ou envoyez-nous vos coordonnées pour une consultation gratuite et sans engagement. Nos ingénieurs commerciaux vous contacteront dans un délai maximum de 1 à 2 jours ouvrables.
Téléchargements de produits
Parlez à nos experts
Vous ne trouvez pas exactement ce que vous cherchez ? Nous proposons des capteurs et des systèmes de mesure standard, mais nous pouvons également vous aider à concevoir un capteur sur mesure ou une solution de mesure complète. N'hésitez pas à nous contacter pour obtenir des conseils ou un devis.
Parler à un expertSpécifications techniques
Spécifications | Valeur |
---|---|
Analogue output | 0 - 10 V, ±10 V, max. 1 mA (short-period short-circuit proof) 4 - 20 mA, max. 500 Ω |
Dimension (w x h x d) | 175 x 80 x 57 mm |
Electrical connection | EMC cable gland on internal terminal block |
Enclosure | EMC aluminium die-cast enclosure (IP65) |
Input resistance | >3 MΩ |
Isolating proof voltage input to output | 200 V, higher isolated proof voltage on request |
Limit frequency (-3 db) | 1 kHz, optional up to 30 kHz |
Max. input sensitivity | 100 mV/V at +10 VDC excitation supply |
Non-linearity | ±0.05 % FSO |
Number of measuring channels | 1 (full bridge resistance >300 Ω) |
Power consumption | 4 W max. |
Strain gauge excitation supply | +5 VDC, +10 VDC |
Supply voltage | 10 - 18 VDC, 18 … 30 VDC, electronics protected against reversal voltage |
Temperature, operating | -20 to +50 °C |
Temperature, storage | -20 to +60 °C |