TD-9000T Digital Indicator
Highly functional model that supports two inputs of a load sensor and a displacement sensor
Key features
- Bridge Voltage: 2.5V / 5V / 10V (30mA MAX.)
- Input range: Strain gauge sensor -3.2mV/V~3.2mV/V
- Zero adjustment range: -3.2mV/V~3.2mV/V
- Calibration method: Equivalent Input / TEDS / Actual Load
- Power supply: DC24V (±10%)
- Compliant with RoHS, CE, UL
About the TD-9000T
Equipped with a 4.3 inch large LCD screen on a compact body sticking to the DIN size. Highly functional model that supports two inputs of a load sensor and a displacement sensor, and has a fulfilling comparison/determination functions, as well as a large internal memory capacity enabling direct storage of waveform data.
Features of the TD-9000T
- Constant Judge function - Real-time display of waveform in hold mode
- Multiple hold function - Up to 5 judgment zones can be set
- Band judge function - Any band can be set and judged
- Compact size - 92mm x 92mm DIN size (same as conventional machines)
- Adoption of large - size monitor - 4.3 inch touch panel color graphic LCD
- Data recording - SD card recording (CSV format) or up to 100 waveform
data can be stored in the main unit memory - Convenient function - Equipped with load cell disconnection detection function
- TEDS standards supported - For the TEDS supported transducers, sensitivity
calibration is automatic calibration - Static strain function - Load cell overloading is obvious
Additional features
- Standard equipped with displacement input (voltage, pulse)
- Communication interface: CC-LINK (option), Ethernet/IP (option), RS-232C (standard)
- Standard equipped with D/A output (0 to 10V, 4 to 20mA)
- Display of statistics information
- Multi-language support: Japanese/English/Chinese/Korean
- Compliant with RoHS, CE, UL
Do you need this sensor for your next project?
Request a no-obligation quote or send us your contact details for a free, no-obligation consultation. Our sales engineers will then contact you within a maximum of 1 to 2 working days.
Product downloads
Talk to our experts
Not exactly what you are looking for? We offer standard sensors, but can also help you with sensors adapted for you or a fully customised measurement solution. Contact us for advice or pricing.
Talk to an expertTechnical Specifications
Specification | Value |
---|---|
A/d conversion | Sampling rate 5kHz/20kHz switchable / Resolution 24-bit (binary) |
Accuracy | Linearity Within 0.01%F.S. ±1digit (Input ≥1.0 mV/V) / Zero drift Within 0.5μV/ Gain drift Within 0.005%F.S/ºC (Input ≥1.0 mV/V)ºC (Input conversion value) / |
Bridge voltage | 2.5V / 5V / 10V (30mA MAX.) |
Calibration method | Equivalent Input / TEDS / Actual Load |
D/a converter | Output (Isolated) Current (4-20mA) Load ≥ 350 ohms Voltage (0.1~10V) set in 1V steps, Load ≥ 2 k ohms |
Environment | Temperature 0ºC~40ºC (Operating)/-20ºC~60ºC (Storage) / Humidity 85% RH or less (without condensation) |
Power supply | DC24V (±10%) |
Sensor input range | Strain gauge sensor -3.2mV/V~3.2mV/V |